The University of Leicester and Photek are pursuing a number of complementary detector development projects for multi-channel detection and imaging of single photon events with time resolution of the order of 25 ps. We describe the HiContent detector, an 8×8 multi-anode MCP detector with optical photocathode, which uses ASIC-based, multi-channel preamplifier/discriminator and time-to-digital convertor electronics, providing 64 channel parallel event detection. We present results from the initial detector calibration which indicate an RMS timing jitter of ≊75ps with a sub-optimal experimental setup.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
T.M. Conneely, J.S. Lapington, J. Milnes,