Article ID Journal Published Year Pages File Type
1825837 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2010 8 Pages PDF
Abstract

This article presents a Monte Carlo simulation of the detector energy response in the presence of pileup in a segmented silicon microstrip detector designed for high flux spectral computed tomography with sub-millimeter pixel size. Currents induced on the collection electrode of a pixel segment are explicitly modeled and signals emanating from events in neighboring pixels are superimposed together with electronic noise before the entire pulse train is processed by a model of the readout electronics to obtain the detector energy response function.The article shows how the lower threshold and the time constant of the electronic filters need to be set in order to minimize the detrimental influence of cross talk from neighboring pixel segments, an issue that is aggravated by the sub-millimeter pixel size and the proposed segmented detector design.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , ,