| Article ID | Journal | Published Year | Pages | File Type | 
|---|---|---|---|---|
| 1825868 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2010 | 6 Pages | 
Abstract
												The advent of the microstrip gas chamber (MSGC) has permitted the development of position-sensitive gas avalanche detectors, which combine sub-millimeter spatial resolution with multi-MHz global rate capabilities not attainable with traditional types of gas counters. Based on massively parallel electronic readout, the need to keep down electronic channel costs imposes a simple front-end design without in-built deadtime compensation. In this report we show that insitu calibration of the deadtime of the detector system is straight-forward and reliable. Methods are described for applications to X-ray and neutron detection.
											Keywords
												
											Related Topics
												
													Physical Sciences and Engineering
													Physics and Astronomy
													Instrumentation
												
											Authors
												J.E. Bateman, D.M. Duxbury, 
											