Article ID Journal Published Year Pages File Type
1826303 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2010 7 Pages PDF
Abstract

A high-resolution powder diffractometer (HRPD) was installed on the IEA-R1 research reactor at IPEN-CNEN/SP, São Paulo, SP, Brazil. It is an extensive upgrade of the old IPEN-CNEN/SP multipurpose neutron diffractometer which used a single boron trifluoride (BF3) detector and a flat mosaic single crystal copper monochromator. The main modifications introduced were the installation of a position sensitive detector (PSD) and a focusing silicon monochromator. Placed at a distance of 1600 mm from sample, the PSD spans an angular range of 20° of a diffraction pattern. An extensive powder diffraction pattern can be obtained by collecting data in contiguous 20° segments. The double-focusing perfect single crystal silicon monochromator, installed in a take-off angle of 84°, can be positioned to produce 4 different wavelengths, namely 1.111, 1.399, 1.667 and 2.191 Å (nominal values). Diffraction patterns, obtained with the instrument, have shown a quite good resolution. Curves full width at half maximum (FWHM) vs. scattering angle (2θ) have been measured with powdered alumina using three vanadium sample holders 1/8, 1/4 and 3/8 in. in diameter. Such curves give the resolution that can be expected from the instrument in an interval 2θ from 5° to 130°. In comparison to the former instrument, the new diffractometer has much better resolution and is, per intensity point, ca. 600 times faster in data acquisition.

Keywords
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , ,