Article ID Journal Published Year Pages File Type
1826687 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2011 9 Pages PDF
Abstract

We measured the energy spectra and spatial distributions of the neutron beam of Accurate Neutron-Nucleus Reaction Measurement Instrument (ANNRI) at the Japan Proton Accelerator Research Complex/Materials and Life Science Experimental Facility (J-PARC/MLF). Our research team designed and built ANNRI to measure nuclear data with high precision. The measurements of the neutron beam were performed on three types of beams provided by ANNRI in the neutron energy range from 1.5 meV to 10 keV. The energy spectra show a typical feature of para-hydrogen moderator, and the absolute intensities almost agree with predictions based on both a simulation calculation of the Japan Spallation Neutron Source (JSNS) and a neutron transmission calculation of the beamline. The available neutron intensities at 21.5 m are 7.5×105, 1.6×104, and 1.1×105 n/cm2/s in the energy ranges 1.5–25 meV, 0.9–1.1 eV, and 0.9–1.1 keV, respectively, under the 17.5 kW JSNS operation. The measured spatial distributions of the beams formed by three different collimators are consistent with those expected from the collimator-system design of the beamline. The beam sizes in FWHM are about 29, 14, and 11 mm for the three different beam collimators. The edges of the spatial distributions are relatively sharp, enabling us to measure the nuclear data successfully.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , , , , , , , , , , , , , , , , , , ,