Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1826803 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2010 | 6 Pages |
Abstract
This work reports on the use of three state-of-the-art Monte Carlo codes (MCNPX, PENELOPE, FLUKA) in the efficiency calibration of a Broad-Energy Germanium (BEGe) detector. Initial discrepancies found between the experimental and computational efficiency values are related to the poor knowledge of some physical parameters of the detector (dead-layers, crystal dimensions, etc.). As a consequence, a sensitivity analysis was carried out. Each parameter was systematically analyzed, and an accurate model of the detector was determined. The obtained results are consistent, allowing this model to be used in computational efficiency calibrations of the equipment at stake.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
R. Luís, J. Bento, G. Carvalhal, P. Nogueira, L. Silva, P. Teles, P. Vaz,