Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1826833 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2009 | 4 Pages |
Abstract
Close‐geometry efficiency calibration has been performed for a 25.4 mm×25.4 mm cylindrical LaBr3(Ce) detector using calibrated point sources of 137Cs and 60Co placed on the top of the detector surface. The absolute total detection efficiency and the absolute photo-peak efficiency, obtained with 60Co, are corrected for coincidence-summing effect. The measured efficiency for 137Cs and the corrected efficiencies for 60Co are reproduced by GEANT4 simulations, considering the detector in a realistic geometry. In addition, the simulations also faithfully reproduce the entire measured spectra for both the sources.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
G. Anil Kumar, I. Mazumdar, D.A. Gothe,