Article ID Journal Published Year Pages File Type
1827074 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2010 5 Pages PDF
Abstract

A dual-axis dual-lateral position-sensitive detector has been developed for charged particle detection. This detector consists of a 20×20 mm square double-sided p-on-n silicon structure with a highly uniform resistive junction and ohmic layers with equi-potential channels. This detector was developed for the possible integration into the FAUST array. In-beam experimental results reveal a detector position resolution better than 200 μm with a non-linearity of 0.002% (rms). The energy resolution as determined from a 228Th alpha source is ∼80 keV for the 6.778 MeV alpha.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , , , , , , , , , , , ,