Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1827169 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2010 | 7 Pages |
Abstract
We developed an image-shifting subtraction program and carried out embossed radiography (ER) utilizing single- and dual-energy subtractions. In particular, dual-energy subtraction was carried out to decrease the absorption contrast of unnecessary regions. The contrast resolution of a target region was increased using the subtraction program and a linear-contrast system in a flat panel detector (FPD). The X-ray generator had a 100 μm-focus tube, and the subtractions were performed at tube voltages of 40 and 70 kV, a tube current of 0.50 mA, and an X-ray exposure time of 5.0 s. ER was achieved with cohesion imaging using the FPD with pixel sizes of 48Ã48 μm2, and the shifting dimension of an object in the horizontal and vertical directions ranged from 48 to 96 μm. At a shifting distance of 96 μm, the spatial resolutions in the horizontal and vertical directions measured with a lead test chart were both 83 μm. In ER of animal phantoms, we obtained high-contrast embossed images of fine bones, gadolinium oxide particles in blood vessels, iodine-based microspheres in coronary arteries approximately 100 μm in diameter.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Eiichi Sato, Akihiro Osawa, Hiroshi Matsukiyo, Toshiyuki Enomoto, Manabu Watanabe, Kiyomi Takahashi, Shigehiro Sato, Akira Ogawa, Jun Onagawa,