Article ID Journal Published Year Pages File Type
1827649 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2009 4 Pages PDF
Abstract

A characterization of scintillating fibres with silicon photomultiplier read-out was performed in view of their possible application in fibre tracking detector systems. Such a concept is being considered for the Kaos spectrometer at the Mainz Microtron MAMI and as a time-of-flight start detector for the hypernuclear physics programme at the P¯ANDA experiment of the FAIR project. Results on particle detection efficiency and time resolution are discussed. In summary, the silicon devices are very suitable for the detection of the low light yield from scintillating fibres insofar a trigger scheme is found to cope with the noise rate characteristics.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , ,