Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1827685 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2009 | 4 Pages |
Abstract
The full-energy peak efficiency calibration and the energy resolution measurements of the 2in.×2in.LaBr3LaBr3γγ-ray detector are presented for γγ-ray energies in the 700 keV–17.6 MeV range. Measurements were done using a combination of proton-capture nuclear reactions on Al27, Na23, K39, Li7 and B11 for high-energy γγ-rays, and radioactive sources such as Co60 and Eu152 for the lowest energies. At high energies, two γγ-rays in a cascade from proton resonance capture were employed using Al, Na2WO4Na2WO4, K2SO4K2SO4 and LiBO2LiBO2 targets. The obtained results were compared to the simulations performed using a GEANT4 code.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
M. Ciemała, D. Balabanski, M. Csatlós, J.M. Daugas, G. Georgiev, J. Gulyás, M. Kmiecik, A. Krasznahorkay, S. Lalkovski, A. Lefebvre-Schuhl, R. Lozeva, A. Maj, A. Vitez,