Article ID Journal Published Year Pages File Type
1827753 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2009 5 Pages PDF
Abstract

A modelling approach for a quick and accurate simulation of γ-spectra from volume and shielded sources is described. It is based on the combination of two web-accessible modelling tools, the Easy Monte Carlo (EMC) and the Gamma Spectrum Generator (GSG), which have been recently developed within the European Commission's Nucleonica portal. The high performance of the simulation was achieved by the use of a next-event estimator for the photon flux calculations and an extensive database of the pre-calculated detector responses for the flux-to-pulse height distribution conversion. The paper presents the first results of the experimental validation of this approach performed with a 10% High Purity Germanium (HPGe) detector and shielded/unshielded reference γ-sources.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation