Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1827754 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2009 | 8 Pages |
Abstract
A charged particle detector array consisting of 50 Si-CsI detector telescopes for study of heavy-ion reactions is under construction in BARC-TIFR Pelletron-LINAC facility. Single-sided segmented silicon-pad detectors of 300 μm thickness and 400 mm2 total active area have been fabricated for this purpose and their design parameters along with static and dynamic characteristics are presented. A low-cost solution for implementing the pulse-shape-based particle discrimination technique with the silicon-pad detectors has been demonstrated through in-beam measurements and the results obtained are presented.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Bency John, A. Topkar, R.P. Vind, A.L. Inkar, D.C. Biswas, A. Saxena, Y.K. Gupta, S. Kailas,