Article ID Journal Published Year Pages File Type
1827904 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2009 4 Pages PDF
Abstract

An interlock system is described that allows volatile targets to be prepared in an evaporator, sealed under vacuum, transported to the γγ-array Gammasphere, and inserted into the reaction chamber for experiments, while maintaining a vacuum better than 10-5torr. The performance of the system is demonstrated for lithium metal targets. A comparison is made between lithium targets covered by a gold evaporation to reduce oxidation and open metal targets transferred in the interlock.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , ,