Article ID Journal Published Year Pages File Type
1828463 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2008 4 Pages PDF
Abstract

A CMOS application-specific integrated circuit (ASIC) called the “Timepix” has been used to readout the pulsed charge signal from various microchannel plate (MCP) stacks. The Timepix 256×256 array is a newer version of the Medipix2 ASIC with added features and modes. One of these modes is the “time over threshold” (TOT) mode where the amplitude of a single charge event can be determined per pixel. With a properly sized input charge cloud from an MCP stack, the centroid of the charge cloud can be determined to sub-pixel accuracy, allowing much higher spatial resolution than the 55 μm pixel size. Using this technique, we demonstrate a spatial resolution better than 57 lp/mm for UV photons, limited by the MCP pore spacing. Also, by changing the top MCP in the stack to a neutron-sensitive MCP, we show a spatial resolution to thermal neutrons of less than 35 μm FWHM. This technique can also be applied to alpha particle imaging using silicon diode arrays bump bonded to a Timepix. Allowing the large conversion charge generated by the alpha particle to diffuse over many pixels, we can centroid 5.5 MeV alphas from 241Am to 18 μm FWHM.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , ,