Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1828479 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2008 | 4 Pages |
Abstract
The combination of Bridgman, recrystallization by dissolving in pure water under hydrothermal conditions and travelling molten zone (TMZ) methods were used for TlBr purification. Detectors of 0.5 and 3 mm thickness were produced from a single crystal grown by the TMZ method. The samples had stable spectrometric characteristics from −40 °C to room temperature. A resolution of 3.7% at 60 keV (241Am source) was attained for 3-mm sample at 0 °C. Electrical and spectroscopic properties of the detectors are reported and discussed.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Vasilij Kozlov, Hans Andersson, Vladimir Gostilo, Markku Leskelä, Alan Owens, Mihail Shorohov, Heikki Sipilä,