Article ID Journal Published Year Pages File Type
1828740 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2009 8 Pages PDF
Abstract

We have investigated the effects caused by proton-induced radiation damage on Multi-Pixel Photon Counter (MPPC), a pixelized photon detector developed by Hamamatsu Photonics. The leakage current of irradiated MPPC samples linearly increases with total irradiated doses due to radiation damage, which is not completely recovered even after a year from the irradiation. No significant change has been observed in the gains at least up to 8.0 Gy (9.1×107n/mm2 in 1 MeV neutron equivalent fluence, ΦeqΦeq). The device has completely lost its photon-counting capability due to baseline fluctuations and noise pile-up after 21 Gy irradiation (2.4×108n/mm2 in ΦeqΦeq), which might be problematic for some applications, such as ring-imaging Cherenkov detectors. We have found that the pulse-height resolution has been slightly deteriorated after 42 Gy irradiation (4.8×108n/mm2 in ΦeqΦeq), where the measured sample has been illuminated with a few hundred photons. This effect should be considered in the case of energy-measurement applications.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , , , , , , , ,