Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1828856 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2008 | 4 Pages |
Abstract
In this paper we describe a single photon method to measure the decay time of two kinds of homemade liquid scintillator. We find that the fast and slow decay time constants of these two samples, with 5 cm diameter and 4 cm height, are about ∼4ns and ∼17ns, respectively. These results are checked through a reference measurement by recording the scintillation pulse shapes at the output of a fast response PMT directly by a digital oscilloscope. The time responses of these two samples are as good as those of commercial liquid scintillator.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Wei-Li Zhong, Zu-Hao Li, Chang-Gen Yang, Jun Cao,