Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1828965 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2008 | 5 Pages |
Abstract
A technique enabling in situ measurements of the spectrum of X-ray tubes employed in mammographic screenings is described. The technique involves the use of a commercially available CCD camera and a set of metal foils and is particularly useful to perform a fast evaluation of the spectral properties of a Mo anode mammographic system operating at standard flux levels. A description of the detector calibration procedure is first given, followed by a discussion of the study and choice of an appropriate set of X-ray attenuation foils. Finally, the use of the system for a spectroscopic characterization of a mammographic system is reported.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
L. Labate, T. Levato, M. Galimberti, A. Giulietti, D. Giulietti, M. Sanna, C. Traino, M. Lazzeri, L.A. Gizzi,