Abstract
A prototype RPC detector with a 2-mm readout strip pitch has been built for research into its position resolution. After determining the optimal threshold and working voltage, the efficiency and position resolution of this prototype RPC detector have been studied carefully. Using cosmic-ray muons with different directions and incident positions, we have obtained less than 0.5-mm average statistical errors of position resolution for two different RPC readout structures: 1- and 1.7-mm readout strip widths. The results show that the RPC can also give high position resolution with very little variation from its traditional structure. This will lead to possible applications of the RPC in medical imaging and other domains.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Ye Jin, Cheng Jianping, Yue Qian, Li Yuanjing, Li Jin, Wang Yi,