1829105 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2008 6 Pages PDF
Abstract
A prototype RPC detector with a 2-mm readout strip pitch has been built for research into its position resolution. After determining the optimal threshold and working voltage, the efficiency and position resolution of this prototype RPC detector have been studied carefully. Using cosmic-ray muons with different directions and incident positions, we have obtained less than 0.5-mm average statistical errors of position resolution for two different RPC readout structures: 1- and 1.7-mm readout strip widths. The results show that the RPC can also give high position resolution with very little variation from its traditional structure. This will lead to possible applications of the RPC in medical imaging and other domains.