Article ID Journal Published Year Pages File Type
1829950 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2008 5 Pages PDF
Abstract

For DEcay SPECtroscopy (DESPEC) experiments with implanted radioactive beams a high resolution gamma tracking array is being designed at GSI. Exploiting the imaging capability of the array, a background suppression method is developed based on a separation of events originating from the target and from localized background sources. Monte-Carlo simulations show that the suppression of the background by a factor from 10 up to 100 or even more can be achieved. The Peak/Total ratio can be as well significantly increased. The dependence of the suppression efficiency on the detector properties is discussed.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, ,