Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1830329 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2007 | 6 Pages |
When an intense sub-nanosecond positron pulse impinges upon a target, a pulse of γ-rays is created which can yield information concerning electron–positron pairs just prior to annihilation. Many conventional γ-ray detectors are unable to exploit the timing information contained within such pulses, and we describe here the development of a fast detector that is able to do so. Using a single-crystal PbF2 Cherenkov radiator coupled to a fast photomultiplier tube (PMT), we have produced a detector with a time response of ∼4 ns (primarily determined by the PMT response), as well as a low-efficiency detector with a sub-nanosecond response. Since 511 keV photons produce very little Cherenkov light, the problem of photomultiplier saturation is mitigated and this detector is therefore well suited to single-shot positron annihilation lifetime spectroscopy (SSPALS) measurements.