Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1830443 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2007 | 5 Pages |
Abstract
The calibration of the gantry angle indicator is an important and basic quality assurance (QA) item for the radiotherapy linear accelerator. In this study, we propose a new and practical method, which uses only the digital level, V-film, and general solid phantoms. By taking the star shot only, we can accurately calculate the true gantry angle according to the geometry of the film setup. The results on our machine showed that the gantry angle was shifted by −0.11° compared with the digital indicator, and the standard deviation was within 0.05°. This method can also be used for the simulator. In conclusion, this proposed method could be adopted as an annual QA item for mechanical QA of the accelerator.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Liyun Chang, Sheng-Yow Ho, Yi-Chun Du, Chih-Ming Lin, Tainsong Chen,