Abstract
The gel dosimetry method was found to be capable of addressing complicated issues related to dose measurements particularly in modern sophisticated radiotherapy applications. The Ferrous-sulphate Xylenol-orange and Gelatin (FXG) radiochromic gel dosemeter is one of the systems used for such applications. Some chemical dosemeters show different responses for low- and medium-energy X-rays in comparison with high-energy γ-photons. The energy and dose rate dependence of the FXG dose response was examined. In addition to the detector response, other important dosimetric properties of the system were investigated for different X-ray beam qualities with tube voltages in the range 100-300 kV. An orthovoltage X-ray therapy unit was used to irradiate standard sized samples of FXG from different batches for radiation doses in the range 0-20 Gy.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
,