Article ID Journal Published Year Pages File Type
1830742 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2007 5 Pages PDF
Abstract

In this study, a miniature fiber-optic radiation detector has been developed using a water-equivalent organic scintillator for electron beam therapy dosimetry. Usually, two kinds of light signals such as fluorescent and Cherenkov lights are generated in a fiber-optic radiation detector when a high-energy electron beam is irradiated. The fluorescent light signal is produced in the scintillator and is transmitted through a plastic optical fiber to a remote light-measuring device such as a PMT or a photodiode. The Cherenkov light could be also produced in the plastic optical fiber itself and be detected by a light-measuring device. Therefore, it could cause problems or limit the accuracy of the detection of a fluorescent light signal that is proportional to dose. The objectives of this study are to measure, characterize and eliminate Cherenkov light generated in a plastic optical fiber used as a component of a fiber-optic radiation detector and to detect a real fluorescent light signal from the scintillator. In this study, the intensity of Cherenkov light is measured and characterized as a function of the incident angle of an electron beam from a LINAC, as a function of the electron beam energy, and as a function of electron beam size. Also, a subtraction method using a background optical fiber without a scintillator and an optical discrimination method using optical filters are investigated to remove Cherenkov light.