Article ID Journal Published Year Pages File Type
1830992 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2007 4 Pages PDF
Abstract

We describe the operation of microchannel plates (MCPs) with 3.2μm diameter channels as photon counting detectors of soft X-rays. Gain and temporal resolution measurements are compared with theoretical scaling laws for channel diameter. A minimum pulse width of 264 ps is observed for a two stage multiplier at a total bias voltage of ∼1930V.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation