Article ID Journal Published Year Pages File Type
1831230 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2006 4 Pages PDF
Abstract

We tested the system resolution of the FOS-CsI (Tl) coupled to CMOS APS for mammography of 28 kV p. Through the results, this detector showed 14 cycle/mm at zero MTF. The reason it was showing lower resolution than was expected was analyzed using the characteristics of scintillator and sensor. The estimation method of system resolution using the ratio of dark current and photocurrent was derived. The resolution of the candidate detectors and conceptual design of pixellated 2-D sensor array performance were investigated through this method. One of the candidate detectors showed desirable resolution of over 20 cycle/mm.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , ,