Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1831272 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2007 | 4 Pages |
Abstract
In this work the performance of novel small animal positron/single-photon emission (PET/SPECT) scanner is estimated via Monte Carlo simulation, considering a YAP/LSO phoswich detector. To overcome the differences between PET and SPECT and in order to simplify the design, the system implements most signal processing digitally with programmable devices. The estimated performance of the described setup, expressed in terms of spatial image resolution and sensitivity, is 1.4 mm/0.6% for PET and 2.5 mm/0.025% for SPECT, figures that are comparable with state of the art dedicated scanners.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Pedro Guerra, Jose L. Rubio, Juan E. Ortuño, Georgios Kontaxakis, Maria J. Ledesma, Andres Santos,