Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1831815 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2007 | 6 Pages |
Abstract
The micropattern time projection chamber (TPC) offers a novel method of imaging the tracks of photoelectrons as a means of X-ray polarimetry with the potential to simultaneously achieve large modulation factors and high quantum efficiency. Measurements with a simple prototype micropattern TPC polarimeter, with an 18 mm-atm absorption depth of a mixture of 50% neon and 50% dimethyl ether, result in a modulation factor of 45.0±0.6% with polarized 6.4 keV X-rays. With unpolarized 5.9 keV X-rays, the measured modulation factor is 0.49±0.54%, consistent with zero. The geometry of the TPC polarimeter will enable substantial improvements in quantum efficiency without the loss of modulation.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
J.K. Black, R.G. Baker, P. Deines-Jones, J.E. Hill, K. Jahoda,