Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1832048 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2006 | 9 Pages |
Abstract
We have developed two analytical models to describe the performance of cryogenic microcalorimeters and bolometers. One of the models is suitable to describe Transition Edge Sensor (TES) detectors with an integrated absorber, the other is suitable for detectors with large absorbers. Both models take into account hot-electron decoupling and absorber decoupling. The differential equations describing these models have been solved using block diagram algebra. Each model has produced closed-form solutions for the detector's responsivity, dynamic impedance, and noise equivalent power for phonon noise, Johnson noise, amplifier noise, 1/f1/f noise, and load resistor noise.
Keywords