| Article ID | Journal | Published Year | Pages | File Type |
|---|---|---|---|---|
| 1832099 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2007 | 18 Pages |
Abstract
JRC has developed the Scrap Neutron Multiplicity Counter (SNMC): an advanced neutron multiplicity counter for the verification of inhomogeneous Pu samples, such as scrap material in MOX fuel fabrication plants. The innovative features of this counter with respect to existing ones rely on two aspects: (i) an optimised design based on Monte Carlo calculations in order to select the most appropriate materials, geometry and detector disposition for maximum efficiency and (ii) novel electronics based on digital signal processing (DSP) reducing the system dead time. The paper recalls the design process, the electronics, the construction and assembly of the counter. Then the results of the first experimental tests will be reported. We will show the characterization of the main physical parameters of the counter, the calibration and the verification of a wide variety of plutonium bearing samples available in the PERLA laboratory at JRC Ispra. This will include pure homogeneous samples (Pu dioxide powders, metal Pu, MOX powders and pellets) and some tests on heterogeneous samples representative of scrap material.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Montserrat Marin Ferrer, Paolo Peerani, Marc R. Looman, Luc Dechamp,