Article ID Journal Published Year Pages File Type
1832335 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2006 4 Pages PDF
Abstract

The aim of this study was to evaluate the effect of K-characteristic radiation on the performance of scintillator crystals incorporated in nuclear medicine detectors (LSO, BGO, GSO). K-characteristic radiation is produced within materials of at least one high atomic number element (e.g. Lu, Gd, Bi). This radiation may either be reabsorbed or it may escape the scintillator. In both cases the light emission efficiency of the scintillator may be affected resulting in either spatial or energy resolution degradation. A computational program, based on Monte Carlo methods, was developed in order to simulate the transport of K-characteristic radiation within the most commonly used scintillator materials. Crystal thickness was allowed to vary from 0.5 up to 15 mm. A monoenergetic pencil beam, with energy varying from 0.60 to 0.511 MeV was considered to fall on the center of the crystal surface. The dominant γ-ray interactions (elastic and inelastic scattering and photoelectric absorption) were taken into account in the simulation. Results showed that, depending on crystal thickness, incident photon energy and scintillator's intrinsic properties (L or K-fluorescence yield, effective atomic number and density), the scintillator's emission efficiency may be significantly reduced and affect spatial or energy resolution.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , ,