Article ID Journal Published Year Pages File Type
1832465 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2006 4 Pages PDF
Abstract

Under industrial radiography conditions, we analyzed short-term radiation characteristics of scintillator coupled CMOS APS (hereinafter SC CMOS APS). By means of experimentation, the contribution of the transmitted X-ray through the scintillator to the properties of the CMOS APS and the afterimage, generated in the acquired image even at low dose condition, were investigated. To see the transmitted X-ray effects on the CMOS APS, Fein focus™ X-ray machine, two scintillators of Lanex™ Fine and Regular, and two CMOS APS array of RadEye™ were used under the conditions of 50 kVp/1 mAs and 100 kVp/1 mAs. By measuring the transmitted X-ray on signal and Noise Power Spectrum, we analytically examined the generation mechanism of the afterimage, based on dark signal or dark current increase in the sensor, and explained the afterimage in the SC CMOS APS.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , ,