Article ID Journal Published Year Pages File Type
1832951 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2006 5 Pages PDF
Abstract

The influence of detection threshold on the performance of ideal photon counting sensors is investigated in mammography for several X-ray tube anode/filter combinations and various breast thicknesses. Based on linear X-ray system theory, a method is proposed to take into consideration detector contrast modulation in the domain of X-ray energy as well as effects related to scattered radiation. Breast screening procedures are modelled as radiographic tasks in which low-contrast objects (tumours or microcalcifications) have to be detected in a noisy image background. For ideal photon counting detectors, results at zero spatial frequency show that higher detection thresholds must be implemented when imaging a thick compressed breast, compared to a thin compressed breast. In practice, the photon counting threshold also affects the image noise and the spatial resolution of hybrid pixel detectors, by tuning the amount of charge-shared events. To be generalized to more realistic imaging situations, this analysis must be performed over the full range of spatial frequencies and include additional detector noise parameters.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , ,