Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1833331 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2006 | 7 Pages |
Abstract
Zero discrimination level setting of liquid scintillation systems for counting of all single photoelectron pulses is required by both triple-to-double coincidence ratio (TDCR) and CIEMAT/NIST methods. However, this requirement may not be able to be met in some circumstances. The influence of rejection of a fraction of the single photoelectron peak has been studied both theoretically and experimentally on 3H, 63Ni, 14C and 90Sr/90Y for both the TDCR and CIEMAT/NIST methods. A modified formula for calculation of efficiency is given in consideration of the rejected fraction of the single photoelectron peak.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
L. Mo, P. Cassette, C. Baldock,