| Article ID | Journal | Published Year | Pages | File Type |
|---|---|---|---|---|
| 1833538 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2006 | 5 Pages |
In recent years the k0-NAA method has been applied and developed at the 15 MW Es-Salam research reactor, which includes: (1) the detection efficiency calibration of γ-spectrometer used in k0-NAA, (2) the determination of reactor neutron spectrum parameters such as α and f factors in the irradiation channel, and (3) the validation of the developed k0-NAA procedure by analysing SRM, namely AIEA-Soil7 and CRM, namely IGGE-GSV4.The analysis results obtained by k0-NAA with 27 elements of Soil-7 standard and 14 elements of GSV-4 standard were compared with certified values. The analysis results showed that the deviations between experimental and certified values were mostly less than 10%.The k0-NAA procedure established at Es-Salam research reactor has been regarded as a reliable standardization method of NAA and as available for practical applications.