Article ID Journal Published Year Pages File Type
185609 Electrochimica Acta 2014 7 Pages PDF
Abstract

An electrochemical lead ion sensor has been developed by modification of carbon paste electrode (CPE) using polypyrrole functionalized with iminodiacetic acid (IDA-PPy) containing carboxyl group. The electrochemical response of Pb2+ ion on the IDA-PPy modified CPE has been evaluated and the controling parameters have been optimized using differential pulse anodic stripping voltammetry (DPASV). The IDA-PPy modified CPE shows a linear correlation for Pb2+ concentrations in the range of 1 × 10−6 to 5 × 10−9 M and the lower detection limit of Pb2+ has been found to be 9.6 × 10−9 M concentration. Other tested metal ions, namely Cu2+, Cd2+, Co2+, Hg2+, Ni2+ and Zn2+, do not exhibit any voltammetric stripping response below 1 × 10−7 M concentration. However, the Pb2+ response is affected in the presence of molar equivalents or higher concentrations of Cu2+, Cd2+ and Co2+ ions in binary systems with Pb2+, consequent to their ability to bind with iminodiacetic acid, while Hg2+, Ni2+ and Zn2+ do not interfere at all. A good correlation has been observed between the lead concentrations as analyzed by DPASV using IDA-PPy modified CPE and atomic absorption spectrophotometry for a lead containing industrial effluent sample.

Graphical abstractFigure optionsDownload full-size imageDownload as PowerPoint slide

Related Topics
Physical Sciences and Engineering Chemical Engineering Chemical Engineering (General)
Authors
, , , , , ,