Abstract

The paper shows the results of a comparison of a developed transported laser interferometer (TLI) with a measurement interferometer XL-80 Renishaw at the distance 0-60 meters. Testings of a breadboard model of the TLI showed that a difference between the travel measurements of the two interferometers does not exceed 6 μm. The mean value of the difference of indications between the TLI and a Renishaw travel measurer at the distance near 58 m approximately equals to 0,5 μm. Root-mean square deviation of the indications of the interferometers approximately equals to 3 μm. At comparison of the sections with the same name between the TLI and the Renishaw travel measurer, measured at different days, a repeatability of the results for the sections with the same name is noted.