Article ID Journal Published Year Pages File Type
1875711 Applied Radiation and Isotopes 2016 4 Pages PDF
Abstract
To calibrate measurement devices for monitoring the activity concentration of 222Rn in air, a prototype of a calibration facility is tested using a solid 226Ra source and a high-purity germanium (HPGe) detector. An emanation box was mounted on the detector for online gamma measurements. Inside this box, a 32.8 kBq ±3% 226Ra standard source was placed. An AlphaGUARD control radon monitor was connected to the emanation box with a pumping air system in an open flow mode as a reference monitor. The emanation coefficient of the source was controlled online by comparing the gamma activity of 214Bi (Eγ=609.3 keV), progeny of 226Ra, to that of the calibration source. A standard 137Cs source, installed within the emanation box, was used as a reference for gamma spectroscopy using the HPGe detector, with a total systematic error of 4% and a random error less than 2%. The ratio between gamma measurements and AlphaGUARD was 0.94±0.4; which is within the 9% uncertainties of AlphaGUARD calibration.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , ,