Article ID Journal Published Year Pages File Type
1876026 Applied Radiation and Isotopes 2014 5 Pages PDF
Abstract
The detection efficiencies of tritium samples by using liquid scintillation counter with hydrophilic plastic scintillator (PS) was approximately 48% when the sample of 20 μL was held between 2 PS sheets treated by plasma. The activity and count rates showed a good relationship between 400 Bq to 410 KBq mL−1. The calculated detection limit of 2 min measurement by the PS was 13 Bq mL−1 when a confidence was 95%. The plasma method for PS produces no radioactive waste.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , , , , , , ,