Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1876026 | Applied Radiation and Isotopes | 2014 | 5 Pages |
Abstract
The detection efficiencies of tritium samples by using liquid scintillation counter with hydrophilic plastic scintillator (PS) was approximately 48% when the sample of 20 μL was held between 2 PS sheets treated by plasma. The activity and count rates showed a good relationship between 400 Bq to 410 KBq mLâ1. The calculated detection limit of 2 min measurement by the PS was 13 Bq mLâ1 when a confidence was 95%. The plasma method for PS produces no radioactive waste.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
Etsuko Furuta, Ryu-ichiro Ohyama, Shigeaki Yokota, Toshiya Nakajo, Yuka Yamada, Takao Kawano, Tatsuhiko Uda, Yasuo Watanabe,