Article ID Journal Published Year Pages File Type
1876052 Applied Radiation and Isotopes 2012 10 Pages PDF
Abstract

Review is made of dosimetric studies of Ge-doped SiO2 telecommunication fibre as a 1-D thermoluminescence (TL) system for therapeutic applications. To-date, the response of these fibres has been investigated for UV sources, superficial X-ray beam therapy facilities, a synchrotron microbeam facility, electron linear accelerators, protons, neutrons and alpha particles, covering the energy range from a few eV to several MeV. Dosimetric characteristics include, reproducibility, fading, dose response, reciprocity between TL yield and dose-rate and energy dependence. The fibres produce a flat response to fixed photon and electron doses to within better than 3% of the mean TL distribution. Irradiated Ge-doped SiO2 optical fibres show limited signal fading, with an average loss of TL signal of ∼0.4% per day. In terms of dose response, Ge-doped SiO2 optical fibres have been shown to provide linearity to x and electron doses, from a fraction of 1 Gy up to 2 kGy. The dosimeters have also been used in measuring photoelectron generation from iodinated contrast media; TL yields being some 60% greater in the presence of iodine than in its absence. The review is accompanied by previously unpublished data.

► We review our TLD studies of Ge-doped SiO2 fibres for therapeutic applications. ► Sources include UV, X-rays, protons, neutrons and alpha particles. ► These TLDs have also been used to measure photoelectrons from I2 contrast media. ► The review is accompanied by previously unpublished data.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , , , , , ,