Abstract
												⺠Dosimetric study of thermoluminescent detectors of CaSO4:Dy, LiF:Mg,Ti and μLiF:Mg,Ti. ⺠Clinical (6 and 15 MV) photon beams dosimetry using liquid water and PMMA phantom. ⺠Linear behavior to the dose range (0.1 to 5 Gy). ⺠TL response reproducibility better than ±4.34%. ⺠CaSO4:Dy represent a cheaper alternative to the TLD-100.
											Related Topics
												
													Physical Sciences and Engineering
													Physics and Astronomy
													Radiation
												
											Authors
												Luciana C. Matsushima, Glauco R. Veneziani, Roberto K. Sakuraba, José C. da Cruz, LetÃcia L. Campos, 
											