Article ID Journal Published Year Pages File Type
1876469 Applied Radiation and Isotopes 2011 6 Pages PDF
Abstract
The modified truncated singular value decomposition (MTSVD) unfolding method is applied to obtain primary spectra for X-ray tubes in radiodiagnostic. Three parameters - voltage, anode angle and filter thickness - of the tube are tested. Unfolded spectra are compared with theoretical extracted from IPEM-78 catalogue. A 2σ error criterion is applied to assess the minimum variations in tested parameters that permits distinguishing between close spectra.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation