Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1876756 | Applied Radiation and Isotopes | 2009 | 5 Pages |
Abstract
Monte Carlo simulations were used to optimize the geometry of a mammography anti-scatter linear grid to achieve minimum scatter-to-primary ratio (SPR) for different X-ray tube voltages. A single optimum design of the grid with 0.9 mm septa height, 12 μm septa thickness and 100 μm interspace thickness was found for breast phantom thicknesses between 30 and 80 mm. The optimal grid has 0.153–0.330 scatter-to-primary ratio, a Bucky factor (BF) less than 2.5 and a contrast improvement factor (CIF) of 1.3.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
M.A. Al Kafi, N. Maalej, A.A. Naqvi,