Article ID Journal Published Year Pages File Type
1876790 Applied Radiation and Isotopes 2009 6 Pages PDF
Abstract

Recent studies on flattening filter (FF) free beams have shown increased dose rate and less out-of-field dose for unflattened photon beams. On the other hand, changes in contamination electrons and neutron spectra produced through photon (E>10 MV) interactions with linac components have not been completely studied for FF free beams. The objective of this study was to investigate the effect of removing FF on contamination electron and neutron spectra for an 18-MV photon beam using Monte Carlo (MC) method. The 18-MV photon beam of Elekta SL-25 linac was simulated using MCNPX MC code. The photon, electron and neutron spectra at a distance of 100 cm from target and on the central axis of beam were scored for 10×10 and 30×30 cm2 fields. Our results showed increase in contamination electron fluence (normalized to photon fluence) up to 1.6 times for FF free beam, which causes more skin dose for patients. Neuron fluence reduction of 54% was observed for unflattened beams. Our study confirmed the previous measurement results, which showed neutron dose reduction for unflattened beams. This feature can lead to less neutron dose for patients treated with unflattened high-energy photon beams.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
,