Article ID Journal Published Year Pages File Type
1876791 Applied Radiation and Isotopes 2009 9 Pages PDF
Abstract

The construction of the adapted patient equivalent phantom (APEP) to simulate the X-ray scattering and absorption by chest, skull and pelvis of nonstandard patient in conventional radiographic equipment is presented. This APEP system is associated to the pre-existing realistic-analytic phantom (RAP) [Pina, D.R., Duarte, S.B., Ghilardi Netto, T., Trad, C. S., Brochi, M.A.C., Oliveira, S.C. de, 2004. Optimization of standard patient radiographic images for chest, skull and pelvis exams in conventional X-ray equipment. Phys. Med. Biol. 49, N215–N226] forming the coupled phantom (RAP–APEP), which is used to establish an optimization process of radiographic images of chest, skull and pelvis for nonstandard patients. A chart of the optimized radiographic technique is established covering a wide range of nonstandard patient thickness, and offering a dose reduction in comparison with those techniques currently used. Different validation processes were applied to confirm the improving of the radiographic image quality when techniques of the established chart are used.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , , ,