Article ID Journal Published Year Pages File Type
1877362 Applied Radiation and Isotopes 2007 10 Pages PDF
Abstract

Backscatter imaging is useful for inspecting structures that are accessible only from one side. However, indications provided by scattered radiation are typically weak, convoluted and difficult to interpret. This paper explores the use of the coded aperture technique to detect flaws using gamma-ray backscatter imaging. The viability of this approach is demonstrated with indications obtained from Monte Carlo simulations of radiation scattering measurements. The results show that, with a 2 mm wide beam of 137Cs photons, flaws as small as 1.5 mm in width can be detected using this technique. Indications of changes in flaw size, location, multiplicity and density were also observable. In addition, it is possible to quantify, from the decoded indications, the flaw location and its size.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, ,