Article ID Journal Published Year Pages File Type
1877565 Applied Radiation and Isotopes 2014 7 Pages PDF
Abstract

•Novel Ge detector having large solid angle for γ-ray measurements was developed.•Correction for coincidence summing was performed with measurements and simulation.•Peak efficiency was determined between 0.05 MeV and 3.2 MeV.

Full-energy peak efficiency at the center position of a through-hole-type clover detector was determined by the measurement of standard sources and by Monte Carlo simulation. The coincidence summing under the large-solid-angle condition was corrected using Monte Carlo calculation based on the specific decay scheme for 133Ba, 152,154Eu, and 56Co. This allowed the peak efficiency to be extended from 0.05 MeV to 3.2 MeV with an approximate uncertainty of 3%.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , , ,