Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1877615 | Applied Radiation and Isotopes | 2014 | 4 Pages |
Abstract
Dose distribution of a High Dose Rate Brachytherapy (BHDR) oncological treatment with 192Ir was measured using a Gafchromic EBT2 film. The film calibration was performed with a 60Co unit and a LINAC of 6Â mV and 18Â mV. Gafchromic behavior of a dosimeter varies in respect of energy. Experimental results of dose distribution match with those planned in the PLATO commercial system, they also show that there is a difference of 2.11% between the planning system and isodoses measured.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
L. Hernández-Ruiz, J.O. Hernández-Oviedo, D. Ruesga-Vazquez, T. Rivera-Montalvo,