Article ID Journal Published Year Pages File Type
187786 Electrochimica Acta 2012 10 Pages PDF
Abstract

This paper reports a transparent Ti microelectrode array (MEA) system for a high-throughput evaluation of bioresponsibility using electrochemical impedance spectroscopy (EIS). The MEA chip integrated with hydroxyapatite (HA) and Ag coatings was selectively prepared by electrochemical deposition based on a novel procedure of multichannel current control. The EIS measurement of living MG63 osteosarcoma cells in the integrated MEA chip was conducted, and the result was analyzed using an equivalent circuit corresponding to a titanium oxide film, protein adsorption layer, cell adhesion layer, and medium. It is shown that the bioresponsibility of Ti–Ag–HA on the MEA chip can be improved, compared with the Ti, Ti–HA, and Ti–Ag coatings. The system was further used for real-time EIS monitoring during continuous cell culture for a long period (12 days). The effect of the long-term cell proliferation on the EIS behavior was discussed. This integrated system is valuable to significantly simplify the operation procedures and quickly evaluate the bioresponsibility of biomaterials.

Graphical abstractFigure optionsDownload full-size imageDownload as PowerPoint slideHighlights► The EIS of living MG63 cells on the Ti MEA chip with Ag, HA, and Ag–HA was monitored. ► The Rcell can be related to the bioresponsibility of the coatings. ► The bioactivity order was evaluated as follows: Ti–Ag–HA > Ti–HA ≈ Ti–Ag > Ti.

Related Topics
Physical Sciences and Engineering Chemical Engineering Chemical Engineering (General)
Authors
, , , , , ,